Loading

ATOVAC

Product

 
home Product MFC MFC

MFC

AFC500/AFC600 MASS FLOW CONTROLLER

AFC500/AFC600 MASS FLOW CONTROLLER

AFC500/AFC600 Mass Flow Controller the mass flow rate of a gas and controls the flow rate according to a given set point. The control range is from 2 to 100% of Full Scale with an accuracy of +/- 1% of Full Scale.

AFC 500 Mass Flow Controller is  available in ranges is from 10 to 50,000 sccm (N2 equivalent)

AFC 500 Mass Flow Controller can interface to complementary ATOVAC equipment(Type GMC1200, GMC220, GMC300D , GMC50) to display the regarding and to provide the power, and set point commands

 

 

Features & Benefits

 Full Scale flow range from 10 sccm to 50 slm for precise and repeatable flow measurement and control

 Percent of full scale accuracy for analog configurations

 Percent of reading accuracy with digital configurations.

 Fast warm-up time minimizes expensive production downtime

 CE Mark compliant


Product



Dimensions








Specifications

Full S cale Ranges (N2 equivalent)

10 SCCM ~ 50 SLM

Maximum lnlet P ressure

150 psig

Normal Operating Pressure  Differential

(with atmospheric pressure at the. MFCoutlet)

10 to 5000 sccm

10000 to 20000 sccm

 

 

10 to 40 psid

15 to 40 psid

Control Range

2% to 100% of F.S 

Accuracy (analog)

(including non-linearity, hysteresis,

and non-repeata bility referenced to 760mmhg and 0°c)

±1.0% of F.S

 

 

 

Repeatability

±0.2% of F.S

Resolution

1.0% of F.S 

Temperature Coefficients

Zero

Span

<0.05 % of F.S./°C

<0.05 % of F.S.

Warm-up Time (to within 0.2% of F.S. of steady state performance)

< 2 min

Controller Settling Time

< 2 sec

Pressure Coefficient

<0.02 % of Rdg./psi

Normal Operating Temperature Range

0°C to 50°C 

Input Voltage Required

Max. current at start-up (first 2 sec)

Typical current at seady state

 

±15 VDC(±5%) @ 200 mA

+15 ~ 24 VDC(±5%) @ 300 mA (Option)

Set Point Command Signal

0 to 5 VDC (Option 4~20 mA)

Output Signal

0 to 5 VDC (Option 4~20 mA)

Connector Types Analog

9-pin "D" Type 

 REMOTE MODERS485, RS232 

Wetted Materials

Stan dard Optional

(seals and valve seat)

316L S.S. , Vition , Kalrez(AFC600)

Leak Integrity    External (scc/sec He)

                          Through closed valve

<1 x 10(-9)

<1.0% of F.S. at 40 psig inlet to atmosphere(Toassure no flow-through, a separate positives shut-off valve is recommended.)

Fittings (compatible with)

Swagelok 4 VCR, 1/4" Swagelok